| Laws
you should know about in electronics and other tips for building and
testing electronic circuits
Ohm's law
The basic law of current flow is Ohm's
law, named for its discoverer, the German physicist Georg Ohm. Ohm's
law states that the amount of current flowing in a circuit made up of
pure resistances is directly proportional to the electromotive force
impressed on the circuit and inversely proportional to the total
resistance of the circuit.
The law is usually expressed by the formula I = V / R, where I is the
current in amperes, V or ( E ) is the electromotive force in volts,
and R is the resistance in ohms (see Electrical Units). Ohm's law
applies to all electric circuits for both direct current (DC) and
alternating current (AC), but additional principles must be invoked
for the analysis of complex circuits and for AC circuits also
involving inductances and capacitances.
Now take a look to a simple circuit :
 |
|
fig.1 - Ohm's
Law |
A potential difference of 10 V (volts)
is needed to make a current of 2A (amps) flow through a conductor with
a resistance of 5 ohms.
Kirchhoff's Laws
If a circuit has a number of
interconnected branches, two other laws are applied in order to find
the current flowing in the various branches. These laws, discovered by
the German physicist Gustav Robert Kirchhoff, are known as Kirchhoff's
laws of networks.
First Kirchhoff Law
 |
|
fig.2 - The
first Kirchhoff Law |
The first of Kirchhoff's laws states
that at any junction in a circuit through which a steady current is
flowing, the sum of the currents flowing to the point is equal to the
sum of the currents flowing away from that point.
Now take a look to the equation in each
point of the circuit :
- A: I2 =I1 + I3
- B: I6 = I2 + I4
- C: I6 = I 3 + I 5
- D: I5 = I1 + I4
Second Kirchhoff Law
 |
|
fig.3 - The
second Kirchhoff Law |
The second law states that, starting at
any point in a network and following any closed path back to the
starting point, the net sum of the electromotive forces encountered
will be equal to the net sum of the products of the resistances
encountered and the currents flowing through them. This second law is
simply an extension of Ohm's law.
Now take a look to the equation for
each path :
Starting B point :
Starting D point :
Starting A point:
Measuring Voltage
 |
|
fig.4 -
Measuring Voltage |
The instrument most generally used to
measure potential difference, or voltage, is a galvanometer with a
high resistance in series with the coil. When such a meter is
connected across a battery, or to two points in an electrical circuit
between which a potential difference exists, a small current (limited
by the series resistor) will pass through the meter. The current is
proportional to the voltage, and the latter quantity can be measured
if the galvanometer is calibrated appropriately. By using the proper
values of series resistors, one galvanometer can be used to measure a
large range of voltages. The most accurate instrument for the
determination of voltage, resistance, or direct current is the
potentiometer, which indicates an unknown electromotive force by
comparing it with a known value.
For the measurement of AC voltages, AC
meters having high internal resistance, or similar meters with high
series resistance, are employed.
Other methods for measuring the value of voltages depend on vacuum
tubes and electronic circuits (see Electronics) and are especially
useful in measurements at high frequencies. One such device is the
vacuum-tube voltmeter. In the simplest form of this meter an AC
voltage is rectified by a diode tube, and the rectified current is
measured by an ordinary galvanometer. Other such voltmeters employ the
amplifying characteristics of electronic tubes to measure extremely
small voltages. The cathode-ray oscilloscope can also be used for
voltage measurements because the deflection of the electron beam is
proportional to the voltage impressed on the deflection plates or
coils.
Measuring Currents
 |
|
fig.5 -
Measuring Current |
Galvanometers are the main instruments
used to detect and measure current. They depend on the fact that force
is generated by an electric current flowing in a magnetic field. The
mechanism of the galvanometer is so arranged that a small permanent
magnet or electromagnet sets up a magnetic field that generates a
force when current flows in a wire coil adjacent to the magnet. Either
the magnet or the adjacent coil may be movable. The force deflects the
movable member by an amount proportional to the strength of the
current. The movable member may have a pointer or some other device to
enable the amount of deflection to be read on a calibrated scale.
In the D'Arsonval galvanometer, a small
mirror attached to a movable coil reflects a beam of light on a scale
about 1 m (about 3 ft) away from the instrument. This arrangement
involves less inertia and friction than does a pointer, and
consequently, greater accuracy is achieved. The instrument is named
after the French biologist and physicist Jacques d'Arsonval, who
devised the first reflecting galvanometer. He also conducted
experiments with the mechanical equivalent of heat and in the
high-frequency oscillating current of low voltage and high amperage,
D'Arsonval current, used in the treatment of certain diseases
(diathermy treatment). The addition of a scale and proper calibration
converts a galvanometer into an ammeter, the instrument used for
measuring electric current in amperes; D'Arsonval was also responsible
for inventing a direct-current (DC) ammeter.
Only a limited amount of current can be passed through the fine wire
of a galvanometer coil. When large currents must be measured, a shunt
of low resistance is attached across the terminals of the meter. Most
of the current is bypassed through this shunt resistance, but the
small current flowing through the meter is still proportional to the
total current. By taking advantage of this proportionality, a
galvanometer can be used to measure currents of hundreds of amperes.
Galvanometers are usually named according to the magnitude of the
currents they will measure. A microammeter is calibrated in millionths
of an ampere and a milliammeter in thousandths of an ampere.
Ordinary galvanometers cannot be used for the measurement of an
alternating current (AC), because the alternation of the current would
produce deflection in both directions. An adaptation of the
galvanometer, however, called an electrodynamometer, can be used to
measure alternating currents by means of electromagnetic deflection.
In this meter a fixed coil, in series with the moving coil, is
employed in place of the permanent magnet of the galvanometer. Because
the current in the fixed and moving coils reverses at the same
instant, the deflection of the moving coil is always in the same
direction, and the meter gives a constant current reading. Meters of
this type can also be used to measure direct currents. Another form of
electromagnetic meter is the iron-vane meter or soft-iron meter. In
this device two vanes of soft iron, one fixed and one pivoted, are
placed between the poles of a long, cylindrical coil through which is
passed the current to be measured. The current induces magnetism in
the two vanes, causing the same deflection no matter what the
direction of the current. The amount of the current is ascertained by
measuring the deflection of the moving vane.
Meters that depend on the heating effect of an electric current are
used to measure alternating current of high frequency. In thermocouple
meters the current passes through a fine wire that heats a
thermocouple junction; the electricity generated by the thermocouple
is measured by an ordinary galvanometer. In hot-wire meters the
current passes through a thin wire that heats and stretches. This wire
is mechanically linked to a pointer that moves over a scale calibrated
in terms of current.
Direct Measuring
 |
|
fig.6 - Digital
Multimeter |
We can also measure voltage or current
by using a digital meter.
There are a few rules in measuring with
digital meters :
1.First move the switch to the area
that specifies what do you want to measure.
2.Then choose a measuring range (Ex: If you are trying to measure 220
V AC first move the switch to the alternative voltage area ,
then in this area move the switch to a number greater then the voltage
you want to measure (in your case you can try 250 ).
3.The next step is the connection of the digital meter to the circuit.
(rules about how you connect the digital meter are specified above ).
4.Last thing that you have to do is to read the digital screen when
the number displayed becomes steady.
Useful links
Resistor
| Capacitor
| Media gallery
|